| |
|
Introduction
Prior to mass spectroscopy proteins in a biological
sample are usually separated by one-dimensional or two-dimensional
gel electrophoresis and made visible by silver- or, better,
(colloidal) Coomassie staining. Each protein on the gel can principally
be identified by MALDI-TOF MS. The analysis involves a number of
steps consecutively consisting of:
-
Spot-picking
-
In-gel digestion and peptides extraction
-
Application of the sample to the target plate
-
MALDI-TOF MS
-
Protein identification
|
Spot-picking
Protein spots can be cut from the gel using a scalpel.
When large numbers of proteins are to be analyzed, it is more convenient
to do this with a robot. An Ettan™ Spot Picker (Amersham Biosciences)
is available at the Nijmegen Proteomics Facility (contact:
L. Lambooy).
|
In-gel
digestion and peptides extraction
In-gel digestion is done by the protocol described
in [8]. The procedure is improved
by the following minor modifications:
- After
the standard washing/dehydration steps, each gel piece (a cylinder
of 1.7 mm3) is re-swelled on ice with 5 ml 12.5 ng/ml trypsin
in 25 mM ammonium bicarbonate for 1 hour.
- Hereafter,
excess trypsin mixture is withdrawn and the gel piece is covered
with 5 ml of 25 mM ammonium bicarbonate containing 0.1% n-Octyl-Pyranoglucoside.
The sample is incubated for 14 hours at 37°C to allow complete
digestion.
-
To extract the peptides, 1 ml 0.5% TFA in 10% acetonitrile is
added to the sample. The mixture is incubated for 1 hour at room
temperature and subsequently sonicated for 2 minutes.
-
A small portion of the extracted fluid (usually 0.2 ml) is directly
taken for MALDI-TOF analysis, i.e. without vacuum-drying.
|
Application
of the sample to the target plate
- Before
MS analysis the extracted peptide mixture is spotted to a metal
target plate together with a matrix solution. Matrix solution
contains 20 mg/ml 4-hydroxy-a-cyanocinnamic acid in 50% acetonitrile/
0.05% TFA.
-
To the target plates used here, 384 samples can be applied. The
plates should be thoroughly cleaned. This is done by consecutively
washing with ethanol, isopropanol, and methanol. To ensure proper
cleaning, each washing step is evaluated under a microscope. If
final washing with methanol is not sufficient, an additional sonication
of the target plate in methanol is performed.
- By
routine, a 0.5 ml of a protein mix is analyzed for external calibration
of the mass spectrometer. The calibration mix contains 1.26mM
bradikinin, 1.26mM P14R, 0.19mM angiotensin and 0.95mM ACTH in
matrix solution. Series of eight samples are spotted around one
external calibration mixture.
- A
critical step in MALDI-TOF MS is the application of the peptide
sample. In literature, the sample is usually pre-mixed with the
matrix, spotted and the mixture is air-dried. The pre-mixing,
however, may result in lower detector signals. Therefore, we prefer
a two-step application. First, 0.2 ml of the peptides extract
is directly spotted to the target plate, directly followed by
the spotting with 0.2 ml matrix solution. The mixture is allowed
to air-dry. In our hands, the two-step procedure results in a
more homogeneous distribution of the protein-matrix crystals and
in higher detector signals.
|
MALDI-TOF
MS
After
protein spotting, the sample plate is transferred into the mass
spectrometer (Bruker Biflex III MALDI-TOF MS; Bruker Daltonics).
The machine is schematically shown in the Figure below, together
with recommended experimental settings.
To
get the best results with respect to accuracy, resolution and
sensitivity, the spectrometer can be operated in different modes
and experimental settings. The settings depend on the nature
of the sample and the finding of the optimal conditions may
be a matter of trial and error. In our analysis of the Methanothermobacter
thermautotrophicus proteome we used the following (see also:
Figure 4):
-
pulsed-ion
extraction (PIE)
-
-
High-voltage- and low-molecular-weight-suppression.
By this setting, low-molecular-weight matrix adducts
are deflected, resulting in the improved signal-to-noise
ratio for peptides <750 Da.
-
Maximum
acceleration voltage of the first electrode: VIS/1=
19 kV. The value is determined as follows: optimal reflection
by the reflectron is achieved at 105% of the applied
acceleration voltage, which is limited at 20 kV. Thus:
VIS/1= 20 kV/1.05 =19 kV.
-
Maximum acceleration voltage of the second electrode:
VIS/2= 9.5 kV, i.e 50% of VIS/1. To obtain a real-time
VIS/2= 9.5 kV, the electrode should be set at 15.6 kV.
-
Detector voltage: 1.9-2.0 kV (setting recommended by
the manufacturer: 1.7 kV). The higher detector voltage
enhances the peak resolution and signal-to-noise ratio.
This is not necessary for high protein concentrations
and, in fact, not very advisable. The detector life
time is reduced and it may lead to a stronger detection
of ghost peaks (false peaks caused by impurities or
matrix adducts) and, eventually, to detector saturation
by strong intensity peaks. In its turn, detector saturation
may result in the loss of weak peptide signals, which
escape detection.
-
Peak shapes are empirically optimized by changing the
lens kV setting (recommended: 9.1 kV).
-
Laser intensity detenuation: 50-55. Higher laser intensity
(>55) may cause peptide fragmentation, whereas lower
intensity (<50) results in lower detection signals.
|
|
Figure
4 - Maldi parameters (click to enlarge) |
| The
MALDI-TOF mass spectrum is generated by mathematically averaging the
spectra obtained from multiple N2-laser shots. In practice, 120-180
shots should be sufficient. The exact position, where the laser hits
the spotted sample is seen on a monitor connected to a microscope
which is focussed on the sample. Inspection through the microscope
also gives an impression of the quality and distribution of the protein-matrix
crystals. It is advisable to collect spectra from different positions
within the sample. |
| |
|
Protein identification
- All
(relevant) peaks seen on the MALDITOF MS spectrum are manually
annotated as their mono-isotopic peptide peaks ( [M+H]1+ ). The
peaks with their respective m/z values are stored in a peak search
list using the X-TOF program (Bruker Daltonics).
-
The search list representing the protein peptide mass fingerprint
is compared to genomic databases. This is done by the MASCOT search
algorithm (MATRIX SCIENCE). MASCOT search uses a peptide mass
mapping search algorithm, which compares experimentally obtained
m/z values with theoretically calculated m/z values of tryptic
peptides of proteins from a genomic open reading frame database.
-
Searching is done under the conditions of a preset peptide mass
error and a preset number of trypsin missed-cleavages. The preset
parameters signify the statistical confidence of the score. For
instance, when the MASCOT search is done allowing a 0.2-0.3 Da
peptide mass error and one missed cleavage by trypsin, the chance
of a false, coincidental protein identification is less then two
times the standard deviation (P<0.05). In addition, the statistical
significance of a hit is expressed as its ‘MOWSE score’:
the higher this number, the higher is the reliability of the hit.
-
Note: It can be very useful to check the theoretical protein pI’s
and/or molecular masses of the positive hits, i.e. as given in
the genomic database, against the experimentally determined values
obtained by two dimensional electrophoresis. The matching of the
experimental and theoretical values, obviously, improves the reliability
of the score.
|
|
|